Figure 3
Figure 3. Intermediate precision assessment. CV of segments from 5 CBUs (UCB1-UCB5) analyzed over 6 days using 2 operators (Op1 and Op2) and 2 Instruments (In1 and In2). The target specification was a CV ≤20%.

Intermediate precision assessment. CV of segments from 5 CBUs (UCB1-UCB5) analyzed over 6 days using 2 operators (Op1 and Op2) and 2 Instruments (In1 and In2). The target specification was a CV ≤20%.