Table 2.

Precision and sensitivity of intracellular IFN-γ assay and LDA

 Replicate 1 Replicate 2 Replicate 3 Replicate 4 Replicate 5 Average of
the replicates 
Flow cytometric intracellular IFN-γ assay 
  Precision* 3700 3570 3570 3220 3220 3456 
 Sensitivity 
  Concentration 1:2 1:4 1:8 1:16 1:160 
  Results 1840 1470 700 460 40 
 1780 1320 820 380 90 
 2000 1240 830 360 60  
  Average of the triplicates 1870 1340 780 400 60 
 Replicate 1 Replicate 2 Replicate 3 Replicate 4 Replicate 5 Average of
the replicates 
Limiting-dilution analysis 
 Precision 11 4 10 6 1.7 4.4 
 Sensitivity2-153 
  Concentration 1:2 1:4 1:8 1:16 1:160 
  Results 1.2 0.8 ND ND ND 
 Replicate 1 Replicate 2 Replicate 3 Replicate 4 Replicate 5 Average of
the replicates 
Flow cytometric intracellular IFN-γ assay 
  Precision* 3700 3570 3570 3220 3220 3456 
 Sensitivity 
  Concentration 1:2 1:4 1:8 1:16 1:160 
  Results 1840 1470 700 460 40 
 1780 1320 820 380 90 
 2000 1240 830 360 60  
  Average of the triplicates 1870 1340 780 400 60 
 Replicate 1 Replicate 2 Replicate 3 Replicate 4 Replicate 5 Average of
the replicates 
Limiting-dilution analysis 
 Precision 11 4 10 6 1.7 4.4 
 Sensitivity2-153 
  Concentration 1:2 1:4 1:8 1:16 1:160 
  Results 1.2 0.8 ND ND ND 

Five independent samples were analyzed to determine the precision of the intracellular IFN-γ assay and LDA, as described in “Materials and methods.” Sensitivity was determined by analyzing samples after serial dilution of the T cells, as described in “Materials and methods.” Results represent absolute numbers of EBV-specific T cells per 105 T lymphocytes.

ND indicates not detectable.

*

Results of 5 replicates of the undiluted concentration are shown.

The results and average of triplicates for the titrations 1:2 to 1:160 are shown.

Results of 5 replicates of the undiluted concentration are shown.

F2-153

The results for the titrations 1:2 to 1:160 are shown.

Close Modal