Table 1.

Results for CDC QC samples measured by PGIMER in 10 independent analytical runs

Mean, nmol/LSDCV, %Range −2 SDRange +2 SDDifference (PGIMER vs CDC), %
CDC WB low QC CDC 232      
 PGIMER 249 22.2 205 294 
CDC WB medium QC CDC 429      
 PGIMER 431 31.9 367 495 
CDC WB high QC CDC 723      
 PGIMER 741 58.2 624 857 
CDC serum low QC CDC 20.0      
 PGIMER 17.5 1.6 14.4 20.7 -12 
CDC serum high QC CDC 53.4      
 PGIMER 46.1 5.1 11 35.9 56.3 -14 
Mean, nmol/LSDCV, %Range −2 SDRange +2 SDDifference (PGIMER vs CDC), %
CDC WB low QC CDC 232      
 PGIMER 249 22.2 205 294 
CDC WB medium QC CDC 429      
 PGIMER 431 31.9 367 495 
CDC WB high QC CDC 723      
 PGIMER 741 58.2 624 857 
CDC serum low QC CDC 20.0      
 PGIMER 17.5 1.6 14.4 20.7 -12 
CDC serum high QC CDC 53.4      
 PGIMER 46.1 5.1 11 35.9 56.3 -14 